| Technical Specification - (
da Vinci alpha ) |
| Test
Range |
: |
2.5 mm to 10 meter(0.100 in to 400 in) (steel). |
| Velocity |
: |
1000
meters/second to 15000 m/sec(40 in/millisec to 600in/millisec).In
Hot key mode it has preset values. |
| Gain |
: |
100 dB Calibrated Gain adjustable in 0.1, 0.5,1,2,6, or
12 dB step |
| Rectification |
: |
Full
wave,Half wave-ve,Half wave +ve and RF mode. |
| Frequency |
: |
It has tuned amplifier with Four Bands:(a) 0.2 MHz to 1MHz
(b) 0.5 MHz to 4 MHz.(c) 0.8MHz to 8MHz(d) 2MHz to 20MHz |
|
Transmitter |
: |
Transmission pulse
Negative spike.(Pulse Rise Time<10ns) and with selectable high(300Vp) or low
(250Vp) power. |
|
Damping |
: |
Damping high/low is selectable.(High= 45ohms,Low=345ohms) |
|
Freeze/Peak Freeze |
: |
A-Scan Freeze, Peak Freeze,Active Echo dynamic available. |
|
Monitor |
: |
Dual gates adjustable in 1% of Screen width with
positive/Negative logic, Gate Expand,Interface trigger modes. |
|
Gate Expand |
: |
Expand Range to width of the gate. |
|
A-Scan memory |
: |
500 Trace Patterns can be stored(with Note/Detail) which
can be recalled, printed or
transferred to PC via USB port. Unlimited no.of A-Scan can
be directly stored in USB Disk with auto file naming |
| Calibration
Set-up |
: |
50
different calibration set-ups can be Stored and Recalled. |
|
Software |
: |
dvaSoft interface software for transferring A-Scan/B-Scan/T-LOG
from da Vinci alpha to PC is supplied as standard
equipment.it also creates sequential,2D and 3D configured
files for thickness logging |
| Display |
: |
High
brightness Colour TFT LCD Display. Display area
320x240 pixel (117 x88mm). Five different colour and Grid scheme options.
Colour leg facility for angle probe for easy
interpretation of skip distance |
|
Reference A-Scan |
: |
Reference
A-Scan pattern of standard test object can be saved and recalled in the
background for easy comparison during testing. |
| DAC |
: |
Dynamic DAC
curve can be Digitally plotted(Smooth parabolic curve) on screen with
selectable additional offset curves From 0 to 14db in
0.1db selectable steps. DAC curve can be plotted using minimum 2 to maximum 10 points
. |
|
DGS |
: |
Defect size evaluation based on 18
predefined probe data and one custom probe set-up per
memory location. Defect
size is directly displayed in ERS value(Equivalent
Reflector Size). |
|
AWS |
: |
Built-in software for evaluation of defects in accordance
with AWS Standards.(AWAD1.1/D1.5) |
Digital
Readout and Trig. Function |
: |
Thickness/Depth
can be displayed in digital readout when using a normal probe and
Sound path, Surface Distance and Depth of echo signals
of GATEa / GATEb are directly displayed when
angle probe is in use. Measurement point can be selected
to be Peak or Flank. Echo height,ERS value,dB diff of DAC/ DGS
curve to single height, Echo height with respect to DAC in
terms of percentage or in dB can be
measured,T-Minimum,Travel distance can be measured when
encoder is connectd and time of travel during Freerun
B-Scan. |
|
Measurement
Unit |
: |
Metric or British unit of measurement is selectable. |
|
Video
output |
: |
VGA video signal output for connecting to monitor/LCD projector. |
|
PRF |
: |
4Hz to 500Hz,Selection in 10 scalable steps.PRF can go
down up to 4Hz when PRF is selected to 1.. |
| Updating
Rate |
: |
60Hz. |
| Power |
: |
Lithium-Ion
Battery pack10.8VDC,7.8AH,give 8 hours continuous
operation from fully charged battery.Battery with the
charge indicator/fuel gauge indicator.da VINCI ALPHA CAN
ALSO OPERATE ON 6 nos.of . C type dry cells |
| Charger |
: |
Input
Voltage 100 to 240 VAC/50Hz. |
| Temperature |
: |
0
to 55° C. |
| Size/Weight |
: |
170x260x110
mm ( HxWxD )/2.3Kgs. With battery |
| (Modification