| Technical
Specification - ( Einstein-II PLUS ) |
| Test Range |
: |
10mm to 5000mm (0.500 in to 200 in) (in steel). In Hot Key mode it has 13 preset values.
Fine mode adjustable in step of 1mm(0.2 in). |
| Velocity |
: |
1000 m/sec to 9999 m/sec (0.0400 in/µs to 0.4000 in/µs). In Hot Key mode it has 7 preset values.
Fine mode adjustable in step of 1 met/sec (0.0001 in/µs). |
| Delay |
: |
Variable from 0 to 3000mm (0 to100 in). In Hot Key mode5mm(0.2 in).
Fine mode adjustable in step of 0.5mm (0.02 in). |
| Gain |
: |
100 dB calibrated gain adjustable in 0.5, 1, 2, 6, 12 or 20 dB step. |
| Rejection |
: |
0 to 100% FSH with LED indicator. |
| Rectification |
: |
Full-wave rectified with filtering. |
| Frequency |
: |
Broad Band amplifier 0.5MHz to 15 MHz. |
| Test Modes |
: |
Pulse echo and transmit/receive. |
| Pulser |
: |
Transmission pulse Negative spike with selectable high or low power. |
| Damping |
: |
Damping high or low is selectable. |
| Freeze/Peak Freeze |
: |
A- Scan Freeze and Peak Freeze available. In Peak freeze it holds echo dynamic pattern which is useful
during angle beam probing to locate peak signal. |
| Connectors |
: |
BNC and LEMO(Size 1) both are provided. |
| Monitor |
: |
Dual gate adjustable in1%of Screen width with Positive/Negative logic, Gate Expand modes. |
| Gate Expand |
: |
Expands Range to width of the gate. |
| A-Scan memory |
: |
200 Trace Patterns can be stored (with Note/Detail), recalled, printed or transferred to PC via RS-232 serial port. |
| Calibration Set-up |
: |
50 different calibration set-ups can be Stored and Recalled. |
| Software |
: |
EinSoft Plus Interface software for transferring A-Scan from Einstein-II PLUS to PC is supplied. |
| Printer Attachment |
: |
Any PC Printer with serial port can be attached for instant print-out of Trace Pattern(A-Scan with Set-up data). |
| Display |
: |
122x92mm(320x240 pixel) Colour LCD/TFT display for better visibility with option of eleven different colour sets. |
| Full Screen |
: |
By pressing Zoom key A-Scan can be displayed in Full Screen area. |
| Reference A-Scan |
: |
Reference A-Scan pattern of standard test object can saved and recalled in the background for easy comparision during testing. |
| DAC |
: |
DAC curve can be Digitally plotted (Smooth parabolic curve) on screen with -6dB and -14dB DAC curves. DAC curve can be set as flaw monitor gate. |
| Digital Readout
And Trig. Function |
: |
Thickness/Depth can be displayed in digital readout when using a normal probe and Beam path, Surface Distance and Depth are directly displayed when angle probe is in use.Measurement point can be selected to be Peak or Flank. |
| Measurement Unit |
: |
Metric or British unit of measurement is selectable. |
| Video Output |
: |
Composite video signal (PAL) output for connection to a TV. |
| PRF |
: |
High or low PRF is selectable. |
| Update Rate |
: |
50Hz. |
| Power |
: |
Lithium-Ion Battery pack 14.4VDC,4AH, gives 8 hours continuous operation from fully charged battery. |
| Charger |
: |
Input Voltage 100 to 240 VAC. Charge status indicator is provided. |
| Temperature |
: |
0 to 55 C. (For operation) |
| Size |
: |
152 x 255 x 70mm (HxWxD). |
| Weight |
: |
2.3 Kgs. with Battery. |
|