ULTRASONIC FLAW DETECTOR >>  EINSTEIN - II PLUS - SERIES
 
>>
‘Video Signal’ Output for connection to a TV
>>
PEAK FREEZE to detect maximum Echo height
>>
Smooth parabolic DAC curves
>>
Note/Label can be attached with each of A-Scan data
>>
Option of FLANK and PEAK Trigger
>>
Selection of Metric or British Unit
 
Technical Specification - ( Einstein-II PLUS )
Test Range :

10mm to 5000mm (0.500 in to 200 in) (in steel). In Hot Key mode it has 13 preset values. Fine mode adjustable in step of 1mm(0.2 in).

Velocity : 1000 m/sec to 9999 m/sec (0.0400 in/µs to 0.4000 in/µs). In Hot Key mode it has 7 preset values. Fine mode adjustable in step of 1 met/sec (0.0001 in/µs).
Delay :

Variable from 0 to 3000mm (0 to100 in). In Hot Key mode5mm(0.2 in).
Fine mode adjustable in step of 0.5mm (0.02 in).

Gain : 100 dB calibrated gain adjustable in 0.5, 1, 2, 6, 12 or 20 dB step.
Rejection :

0 to 100% FSH with LED indicator.

Rectification :

Full-wave rectified with filtering.

Frequency :

Broad Band amplifier 0.5MHz to 15 MHz.

Test Modes : Pulse echo and transmit/receive.
Pulser : Transmission pulse Negative spike with selectable high or low power.
Damping :

Damping high or low is selectable.

Freeze/Peak Freeze : A- Scan Freeze and Peak Freeze available. In Peak freeze it holds echo dynamic pattern which is useful during angle beam probing to locate peak signal.
Connectors : BNC and LEMO(Size 1) both are provided.
Monitor :

Dual gate adjustable in1%of Screen width with Positive/Negative logic, Gate Expand modes.

Gate Expand :

Expands Range to width of the gate.

A-Scan memory :

200 Trace Patterns can be stored (with Note/Detail), recalled, printed or transferred to PC via RS-232 serial port.

Calibration Set-up :

50 different calibration set-ups can be Stored and Recalled.

Software :

EinSoft Plus Interface software for transferring A-Scan from Einstein-II PLUS to PC is supplied.

Printer Attachment :

Any PC Printer with serial port can be attached for instant print-out of Trace Pattern(A-Scan with Set-up data).

Display :

122x92mm(320x240 pixel) Colour LCD/TFT display for better visibility with option of eleven different colour sets.

Full Screen :

By pressing Zoom key A-Scan can be displayed in Full Screen area.

Reference A-Scan :

Reference A-Scan pattern of standard test object can saved and recalled in the background for easy comparision during testing.

DAC :

DAC curve can be Digitally plotted (Smooth parabolic curve) on screen with -6dB and -14dB DAC curves. DAC curve can be set as flaw monitor gate.

Digital Readout And Trig. Function :

Thickness/Depth can be displayed in digital readout when using a normal probe and Beam path, Surface Distance and Depth are directly displayed when angle probe is in use.Measurement point can be selected to be Peak or Flank.

Measurement Unit :

Metric or British unit of measurement is selectable.

Video Output :

Composite video signal (PAL) output for connection to a TV.

PRF :

High or low PRF is selectable.

Update Rate :

50Hz.

Power :

Lithium-Ion Battery pack 14.4VDC,4AH, gives 8 hours continuous operation from fully charged battery.

Charger :

Input Voltage 100 to 240 VAC. Charge status indicator is provided.

Temperature :

0 to 55 C. (For operation)

Size :

152 x 255 x 70mm (HxWxD).

Weight :

2.3 Kgs. with Battery.