| Test
Range |
: |
10
mm to 5000 mm ( in steel ). In Hot Key Mode it has 13 preset values.
In Fine Mode it is adjustable in step of 1mm. |
| Velocity |
: |
1000
meters/second to 9999 meters/second . In Hot Key Mode it has 7 preset
values. In Fine Mode it is adjustable in step of 1 meter/second, |
| Delay |
: |
Variable
from 0 to 3000 mm. |
| Gain |
: |
Total
Amplifier Gain : 140 dB. Calibrated Gain : 100 dB adjustable in steps
of 0.5,1,2,6,12 or 20 dB. |
| Rejection |
: |
0
to 80% FSH with LED indication . |
| Rectification |
: |
Full
wave rectification with filtering. |
| Frequency |
: |
Broad
Band amplifier 0.5 MHz to 15 MHz. |
| Test
Modes |
: |
Pulse
Echo and Transmit/Receiver. |
| Connectors |
: |
BNC
and LEMO ( Size 1 ) both are provided. |
| Monitor |
: |
Dual
gate adjustable in 1% of Screen width with Positive/Negative logic,
Gate Expand modes. |
| Gate
Expand |
: |
Expands
Range to width of the gate. |
| A-Scan
memory* |
: |
200
Trace Patterns can be stored, recalled, printed or transferred to
PC via RS-232 serial port. |
| Calibration
Set-up |
: |
50
different calibration set-ups can be Stored and Recalled. |
| Software* |
: |
Using
“EinSoft TFT” software A-Scan with calibration set-ups can be transferred
to PC via RS-232 serial link for storage in PC or printing. |
| Printer
Attachment |
: |
IBM compatible Printer having serial port can be directly attached
to main unit for printing of stored A-Scan waveform with calibration
data. |
| Display |
: |
High
brightness Colour TFT LCD Display. Display area 122x92mm (320 x 240
Pixel). Eleven different colour scheme options. |
| Full
Screen |
: |
By
pressing Zoom key A-Scan can be displayed in Full Screen area. |
| Reference
A-Scan* |
: |
Reference
A-Scan pattern of standard test object can saved and recalled in the
background for easy comparision during testing. |
| DAC* |
: |
DAC
curve can be entered using minimum 2 to maximum 10 points and then
digitally drawn with –6dB and –14dB curves. |
Digital
Readout* and
Trig. Function |
: |
Thickness/Depth
can be displayed in digital readout when using a normal probe and
Beam path, Surface Distance and Depth are directly displayed when
angle probe is in use. |
| Update
Rate |
: |
50Hz. |
| Power |
: |
Lithium-Ion
Battery pack 14.4VDC, 4AH, gives 8 hours continuous operation from
fully charged battery. |
| Charger |
: |
Input
Voltage 100 to 240 VAC. Charges status indicator is provided. |
| Temperature |
: |
0
to 55° C. |
| Size |
: |
152x255x70
mm ( HxWxD ) |
| Weight |
: |
2.3
Kgs. with Battery. |
| (Modification
rights reserved) |
(