| Test Range |
: |
0 - 10 mm
Min. to 0 - 5 meters Max. ( in Steel ). 10 mm, 50 mm, 250 mm and 1Met . coarse range steps. Can be continuously varied upto 5x with fine control.
|
| Pulse Shift |
: |
0- 500 mm
( in steel ) |
| Frequency Range |
: |
0.5 to 15
MHz [ Switch able to 0.5 to 4 MHz., 2 to 12 MHz and 0.5 to 15 MHz. (
wide band ) ]
|
| Gain |
: |
Total Gain : 140 dB
Calibrated Gain : 100 dB
Coarse : 0/20/40/60 dB
Fine : 0-40 dB ( 2dB Steps )
Super-Fine : Uncalibrated, continuously variable between 0 to 6 dB |
| Reject |
: |
Unwanted signals can be suppressed/rejected
( Between 0 to 80% FSH ). Unaffected signals remain linear and of the same height. LED warning light when switched on. |
| Variable Pulse Energy |
: |
Can be switched from maximum pulse energy mode to continuously variable
mode ( Variable damping ) for the best possible matching to the respective probe, test piece and application requirement. |
| Flaw Monitor Gate |
: |
Single Channel Monitor with all the controls on front panel. Also selection of positive or negative Logic, visual or
switchable audible alarm. Gate width adjustable from 3 mm to 100%, gate threshold adjustable from 10% to 100% FSH. Gate Width is independent of test range selected. |
| Gate Switching output |
: |
12 VDC, load max 500 mA. |
| Display |
: |
Cathode Ray Tube
( CRT ) display, 70 x 55 ( mm ) size with internally etched graticule. |
| Power |
: |
(a) Battery : Heavy duty (12 Volts / 7 AH) Lead Acid battery, 10-12 hours operating
time between charges. LED Indication for low battery.
(B) Mains : On 230 V AC supply using Batter Charger Cum AC Adapter. |
| Size |
: |
130 x 250 x 330(mm) (H x W x D) |
| Weight |
: |
(a) Without Battery:
(b) With Battery
4.5 Kgs (MSL-32)
7.5 Kgs. (MSL-32)
5.0 Kgs.(MSL-48)
8.0 Kgs. (MSL-48) |
| (
ONLY MSL - 48 ) |
| Electronic DAC |
: |
To correct for amplitude variation due to change in distance caused by energy losses as well as near field and beam spread effects. The shape of curve displayed can be recorded for a particular set-up to duplicate it in future without the need of test blocks. |
| Digital
Thickness / Distance |
: |
The direct digital readout simplifies wall thickness measurement (0.01 mm resolution). Flaw
Depth and Surface Distance can be directly displayed on digital read-out while doing weld- inspection. |